Description:
ICCB-Longwood has two Multidrop Combi nL bulk reagent dispensers capable of filling 96-, 384- and 1536-well microplates. Because the fluidics are so fine, it is prone to clogging and there is a strict cleaning protocol that must be adhered to both before and after use. The Combi nL uses a pressurized reagent bottle, tubing, and valve assembly to dispense (no cartridges). It cannot be removed and autoclaved – ethanol cleaning is the only method for “sterilization”. It is also recommended that mild detergents, such as a 1% Micro-90 solution, can be used for cleaning and preventing clogging.
Best for:
-
Non-sterile low volume dispensing
-
Low priming volume needs
-
1536-well plate applications
Not good for:
-
Cells or protein aggregates which are large or “clump”
-
Applications where sterility is critical.
Location:
SGM 611 bench top
SGM 612 bench top