Thermo Combi nL

 

Thermo Multidrop Combi nL (for smaller volume dispensing)

  • Bulk reagent dispenser capable of filling 96-, 384- and 1536-well microplates. 
  • Volume range of 50 nL to 50 uL, with a low priming volume (1-3 mL). 
  • Dispense any volume, in its range, to any well of a plate using the attached computer – no row or column restrictions. 
  • Requires a strict cleaning protocol that must be adhered to both before and after use. 

Request Training

Schedule Equipment

Description:

ICCB-Longwood has two Multidrop Combi nL bulk reagent dispensers capable of filling 96-, 384- and 1536-well microplates. Because the fluidics are so fine, it is prone to clogging and there is a strict cleaning protocol that must be adhered to both before and after use. The Combi nL uses a pressurized reagent bottle, tubing, and valve assembly to dispense (no cartridges). It cannot be removed and autoclaved – ethanol cleaning is the only method for “sterilization”.  It is also recommended that mild detergents, such as a 1% Micro-90 solution, can be used for cleaning and preventing clogging.

Best for: 
  • Non-sterile low volume dispensing
  • Low priming volume needs
  • 1536-well plate applications
Not good for:
  • Cells or protein aggregates which are large or “clump”
  • Applications where sterility is critical.

Location:

SGM 611 bench top
SGM 612 bench top